What parameters are affecting while generation of Cheyne Stokes.

11 posts
bio
Was this reply useful? Learn more...
   
[-]
skvbksc321 +0 points · over 1 year ago Original Poster

Can anyone guide me please what factors affects to generate cheyne stokes graph. How to generate cheyne stoke graph using ASL 5000. So, the Resmed AIRSENSE 10 AUTOSET detect it as Cheyne Stoke respiration

3,161 posts
bio
Was this reply useful? Learn more...
   
[-]
Sierra +0 points · over 1 year ago Sleep Patron

Is see that the company that makes this breathing simulator, IngMar Medical, makes 35 different standard patient models, but claim they can be easily tweaked based on specific needs. There is a Kussmaul Breathing pattern listed but not the specific Cheyne-Stokes Respiration. But they provide a link to contact them for specific requests.

The ResMed AirSense 10 does detect and report CSR, but it is not always accurate. Have a look at this thread which shows some sleep reports from an AirSense 10 reporting CSR (areas highlighted in green). I believe a standard CSR is waxing and waning breathing effort with central apnea events between them. The events reported in this thread do not have the smooth waxing and waning, and also have obstructive events between each cycle instead of central events. To me a true CSR pattern is a sine wave of smooth breathing effort with CA events between more of them. In other words breathing effort is reduced until there is no flow, and there is no obstruction.

The Phillips DreamStation machines report Periodic Breathing instead of specifying it as CSR. I'm guessing that the reason is that their detection method is really not specific enough to be sure it is CSR. ResMed are reporting it, but perhaps not always correctly.

I am a mechanical engineer, and not a medical professional, but CSR would seem to be easy to simulate. You would just use a slow sine wave function to drive breathing effort. The speed of the wave and time between waves would need to be about 12 seconds or so to allow the apnea machine time to identify a central apnea event. Here is an example from my personal sleep reports of what I believe a true CSR event looks like. In this case the ResMed has identified it correctly as CSR.

Hope that helps some,

11 posts
bio
Was this reply useful? Learn more...
   
[-]
skvbksc321 +0 points · over 1 year ago Original Poster

Hi Sierra,

Can you please provide some more clarity about the topic or else if you have any .vr3 file of cheyne stoke can you share to me. I would also like to know what parameters we should change, so we can be able to generate Cheyne stoke graph.

3,161 posts
bio
Was this reply useful? Learn more...
   
[-]
Sierra +0 points · over 1 year ago Sleep Patron

I am afraid I don't know much about the ASL 5000. I just use a software program like OSCAR to view my data.

11 posts
bio
Was this reply useful? Learn more...
   
[-]
skvbksc321 +0 points · over 1 year ago Original Poster

Ok sierra thank you for your suggestion.

Please be advised that these posts may contain sensitive material or unsolicited medical advice. MyApnea does not endorse the content of these posts. The information provided on this site is not intended nor recommended as a substitute for advice from a health care professional who has evaluated you.