We use cookies and other tools to enhance your experience on our website and to analyze our web traffic.
For more information about these cookies and the data collected, please refer to our Privacy Policy.

jeffez

jeffez
Joined Feb 2019
Bio

Just starting with Resmed Aircurve 10S Feb 2019

Houston

jeffez
Joined Feb 2019
Bio

Just starting with Resmed Aircurve 10S Feb 2019

Houston

Thanks. My take is 167 degrees seems OK for Resmed hoses for "High level thermal disinfection" . (see below from Resmed manual) What about my DreamWear Hybrid Full Mask with Headgear? Their manual says disassemble then "hand wash in warm water with liquid dishwashing detergent, rinse, lay or hang to dry". Also you may wash mask only (not fabric straps-do by hand) once/week on top shelf of dishwasher, no DRY cycle. After my first 3 weeks, I have done none of these cleanings. (My bad!)

Reprocessing the air tubing Disconnecting

  1. Hold the cuff of the air tubing and gently pull it away from the device.
  2. Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart. Decontaminating Before the disinfection process, each component must be cleaned and rinsed so no visible contamination is present.
  3. Clean all components with a soft bristled brush for one minute while soaking in detergent solution (see table below). Pay particular attention to all crevices and cavities.
  4. Run the detergent solution through the air tubing repeatedly until no contamination is visible.
  5. Thoroughly rinse each component according to the detergent manufacturer's instructions. ResMed has tested the following detergents according to the manufacturer’s instructions: Detergent Water temperature SlimLine ClimateLineAir Standard Alconox™ (diluted at 1%) Hot water (approx 140°F or 60°C) Warm water (approx 113 to 140°F or 45 to 60°C) Room temperature water (approx 70°F or 21°C)    Neodisher MediZym™ (diluted at 2.0%) Warm water (approx 113 to 140°F or 45 to 60°C)    High level thermal disinfection Part Validated number of cycles Hot water: 167°F (75°C) for 30 minutes. SlimLine 20 ClimateLineAir 10 Standard 100
  6. Immerse the air tubing in a water bath. Take care that no air bubbles are trapped inside the air tubing.
  7. Increase the water bath temperature to 167ºF (75°C) for 30 minutes. Higher temperatures may damage the tubing.
  8. Air dry out of direct sunlight and/or heat. 24 Inspecting Perform a visual inspection of the air tubing. If any visible deterioration is apparent (holes, tears or cracks etc), the air tubing should be discarded and replaced. Slight discoloration may occur and is acceptable. Packaging and storage Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -4°F to 140°F (-20°C to 60°C).